Loading…

Loading grant details…

Completed NON-SBIR/STTR RPGS NIH (US)

Development of high-throughput, high-sensitivity EPR sample handling capabilities for biomedical research

$3.72M USD

Funder NATIONAL INSTITUTE OF GENERAL MEDICAL SCIENCES
Recipient Organization Medical College of Wisconsin
Country United States
Start Date Jan 01, 2021
End Date Feb 28, 2025
Duration 1,519 days
Number of Grantees 2
Roles Principal Investigator; Co-Investigator
Data Source NIH (US)
Grant ID 10744232
Grant Description

Project Summary/Abstract Electron paramagnetic resonance (EPR) spectroscopy is a critically important technique in biomedical research with a unique ability to detect naturally occurring or engineered unpaired electrons in complex biological environments. EPR has wide-ranging applicability to structural biology, metalloprotein research, redox biology,

rational drug design, and clinical diagnostics. Groundbreaking advancements in sample volume requirements for biomedical EPR spectroscopy applications were made nearly four decades ago with the development of the loop-gap resonator (LGR), which represented breakthrough benefits such as 10-fold lower sample volume

requirements and higher resonator efficiencies to increase signals over the commonly used cavity resonator. To accommodate the biomedical needs of even further increased signal intensity and to provide the broader scientific community with accessible technology, we propose to capitalize on our recent development of the

dielectric LGR (dLGR) concept and optimize this resonator technology to increase sensitivity beyond that achieved upon introduction of the LGR. A dLGR is effectively a small dielectric resonator placed inside the inner loop of an LGR where the return flux of the dielectric flows through the outer loops of the LGR. Analytic theory

and our high-frequency structure simulations indicate that the dLGR enables an order-of-magnitude improvement in sensitivity over the LGR. To take full advantage of the extremely low volume requirements for the dLGR, we propose to develop efficient sample handling technologies that couple the dLGR to high-

throughput sample handling instrumentation. We aim to develop two transformative technologies for biomedical EPR applications: i) a nano-dLGR with a 10-fold increase in sensitivity for ~0.2 µL sample volumes integrated with a customized autosampler and ii) a micro-dLGR for a dramatic increase in sensitivity for ~2 µL sample

volumes with an optimized stopped-flow system for millisecond time scale kinetics measurements. These transformative and innovative prototypes with outstanding sensitivity will be easy to use and ultimately widely available to the scientific community. Where the LGR was a transformative advance compared with cavity

resonators, the dLGR is expected to be another transformative leap from an LGR.

All Grantees

Medical College of Wisconsin

Advertisement
Discover thousands of grant opportunities
Advertisement
Browse Grants on GrantFunds
Interested in applying for this grant?

Complete our application form to express your interest and we'll guide you through the process.

Apply for This Grant